Remove the Shipping Hardware—BIOSTACK2WR | 19
Wrist/gripper cover removed exposing small wrist shipping bracket
4. Remove the gripper cover (with the BioStack label) using the Phillips
screwdriver to expose the small wrist shipping block.
5. Use the 3/32 hex wrench to remove the wrist shipping block.
6. Reinstall the wrist/gripper cover.
7. Store the screws with the other packaging materials in the shipping boxes.