58 | Chapter 3: Operation
Readers
Gen5/Gen5 TS
l Select System > Instrument Control to get the reader's software information.
Precision/XS
BioStack PC Control Software
l Click Get On-board values now.
If software information is displayed without error, the BioStack’s Checksum Test
completed successfully.
If an error code displays, look it up in the Troubleshooting chapter.
l If the System Test fails, the LED status light flashes repeatedly. If this happens, turn
off the BioStack. Make sure there is no plate on the carrier, and check for any
obstructions. Ensure that all of the shipping hardware has been removed. If you cannot
resolve the problem, contact BioTek’s Technical Assistance Center for guidance.
BioStack™ Microplate Stacker