EasyManua.ls Logo

BK Precision MR Series - Current Calibration

BK Precision MR Series
71 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Calibration 61
9.2 Current Calibration
To perform a current calibration:
Connect the shunt resistance, DMM and Electronic load to the output of the power supply as shown in the gure
below:
Figure 9.2 Current Calibration Diagram
Set the DMM to a DC voltage measurement. Set the Electric Load in CV mode, the voltage set 5V. Select the
CURRENT parameter in the CALIBRATION menu and press . The following will be displayed:
The 1/4 signies there are 4 points to be calibrated, and the current calibration point is 1, and press to start
calibration.
When METER parameter is blinking input the current, the voltage reading from the DMM divided by the shunt
resistance, in the METER parameter by using the numeric keypad and pressing .
Repeat this procedure by entering the meter readings for points 2 4.

Table of Contents

Related product manuals