Calibration 62
9.3 OVP Calibration
The overvoltage protection (OVP) calibration can be executed right after the voltage calibration or the user can select
OVP in the Calibration menu.
To perform the overvoltage protection calibration:
1. Disconnect any load from the output terminal of the power supply.
2. Select the OVP parameter in the CAL menu and press . The following will be displayed:
3. Press the to run the OVP calibration or press to abort. The OVP calibration is an automatic process
that will take about 10 to 20 seconds.
4. After the calibration is nished, the following message will be displayed: