Calibration 63
9.4 OCP Calibration
The overcurrent protection (OCP) can be executed right after the current calibration or the user can select OCP in the
Calibration menu.
To perform the voltage calibration:
1. Short the output + and - terminals of the power supply as shown below:
Figure 9.3 OCP Calibration Diagram
2. Set the Electric Load in CV mode, the voltage set 5V.Select the OCP parameter in the CAL menu and press
. The following will be displayed:
3. Press to run the OCP calibration or press to abort. The OCP calibration is an automatic process that
will take about 20 to 30 seconds.
4. After the calibration is nished, the following message will be displayed: