Description of Cable Conduits
# From To Cable
Length
Cable Di-
ameter
Largest Plug Diame-
ter
Minimum Conduit Di-
ameter
A
Computer cabi-
net (1.0)
Generator cabinet 1
(2.0a)
30 m 6 mm 35 mm
50 mm (low voltage
cable)
B1
Computer cabi-
net (1.0)
In-room distribution
box (1.3)
30 m 9, 10 mm No plug
75 mm (low voltage
cable)
B2
1)
Computer cabi-
net (1.0)
In-room distribution
box (1.3)
30 m 5, 6 mm 35 mm
75, 100 mm (low volt-
age cable)
C
1)
Computer cabi-
net (1.0)
Flat panel detectors
(2.3a/2.3b)
30 m
2x 8 mm,
2x 9 mm
52 mm
75, 100 mm (low volt-
age cables)
D1
Computer cabi-
net (1.0)
Control room distri-
bution box (1.5)
30 m 9, 10 mm No plug
75 mm (low voltage
cables)
D2
Computer cabi-
net (1.0)
Control room distri-
bution box (1.5)
30 m 5, 11 mm 53 mm
75 mm (low voltage
cable)
E
2)
Computer cabi-
net (1.0)
Linac gantry stand
(0.0) for V
arian
30 m
6 mm 35 mm
50 mm (low voltage
cable)
G1
In-room distri-
bution box
(1.3)
Optional: Ceiling-
mounted touch
screen (1.2)
8 m
4, 5.9, 2x 7
mm
35 mm
50 mm (low voltage
cable)
G2
In-room distri-
bution Box
(1.3)
Standard: Wall-
mounted touch
screen (1.2)
8 m
5.9, 2x
7 mm
35 mm
50 mm (low voltage
cable)
H1
Computer cabi-
net (1.0)
Camera system (1.1) 30 m 7, 8 mm 35 mm
50 mm (low voltage
cables)
H2
In-room distri-
bution box
(1.3)
Camera system (1.1) n/a 5 mm No plug
25 mm (see page
77)
I
In-room distri-
bution box
(1.3)
Alternatively: PE
grounding spot
n/a 9 mm No plug
25 mm (see page
77)
J
In-room distri-
bution box
(1.3)
External bluetooth
module (3.2)
5 m 6 mm 10 mm
n/a (low voltage ca-
bles)
K
Computer cabi-
net (1.0)
PE grounding spot n/a 9 mm No plug
25 mm (see page
77)
M
3)
Generator cab-
inet 1 (2.0a)
Left side: Floor box 1
(2.1a)
11, 15,
18 mm
9, 12, 17
mm
Diameter: 70 mm Plug
length: 180 mm
130 mm [trench:
100mm] (high voltage
cables, no sharp
turns)
M
3)
Generator cab-
inet 2 (2.0b)
Right side: Floor box
2 (2.1b)
11, 15,
18 m
9, 12, 17
mm
Diameter: 70 mm Plug
length: 180 mm
130 mm [trench:
100mm] (high voltage
cables, no sharp
turns)
O
Computer cabi-
net (1.0)
X-ray warning light 1
& 2 (2.5a/2.5b)
30 m +
30 m
4 mm No plug
25 mm (low voltage
cables)
P
Generator cab-
inet 1 (2.0a)
PE grounding spot n/a 9 mm No plug
25 mm (see page
77)
R
Computer cabi-
net (1.0)
External bluetooth
module (3.2)
30 m 6 mm 35 mm
50 mm (low voltage
cable)
SITE SPECIFICATIONS