EasyManua.ls Logo

Bruker DektakXT - Adjusting X-Y Position with Manual Controls; Leveling the Stage Manually

Bruker DektakXT
83 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-6 Basic Operations and Maintenance
USING THE MANUAL STAGE-POSITIONING CONTROLS
If your DektakXT system includes the manual sample-positioning stage (see Figure 3-8), you must
use the procedures described in this section to adjust both the X and Y positions.
Adjusting the X-Y Position
To position a sample using the manual sample-positioning stage controls:
1 Ensure the stylus is not touching the sample surface. If it is, click the Tower Up button
on the
toolbar.
2 Position the scan start site by using the X-Y positioning levers.
Pull the lever below the front of the platform for coarse X movement.
Pull the lever below the left side of the platform for coarse Y movement.
Use the respective knobs to finely position the sample.
Figure 3-8: X-Y Positioning Levers on the Manual Sample-Positioning Stage
NOTE – After fine-tuning the scan site using the X and Y positioning levers, you may have
to re-null the tower to adjust for skew. (To re-null the tower, tower up and then tower down.)
Leveling the Stage
All stage configurations of the DektakXT system allow you to manually level the sample-positioning
stage by turning the leveling knob below it (see Figure 3-9). The closest possible manual leveling
ensures the best profiler performance. Guidelines for leveling the stage appear after the picture.
X coarse-
positioning Lever
Y coarse-
positioning
lever
X fine-positioning
knob

Table of Contents

Other manuals for Bruker DektakXT

Related product manuals