6 | Operating the interface pro BÜCHI Labortechnik AG
52/88 Operation Manual UniversalExtractor E-800
6.5 Selecting a method for an extraction position
WARNING
Risk of fire or explosion.
The consequences are device damage, serious injuries or death.
u Make sure that the ignition point of the solvent is higher than 160 °C.
NOTE
Combination of solvents with far apart boiling points.
4 For efficient drying do not use solvents with boiling point differences higher than
20 °C.
Navigation path
➔
u Navigate to the [ Home ] menu via the navigation path.
u Tap the extraction position you wish to edit.
u The display shows a dialog box with selectable methods.
u Select a method.
u Tap the function [ OK ] on the function bar.
u Select a method for each extraction position by repeating the previous four steps.
Save the selection with the following process description
Navigation path
➔
u Navigate to the [ Setup ] menu via the navigation path.
u Tap the function [ New ] on the function bar.
ð The new setup is created with a default name.
u Tap the new setup.
u Tap the function [ OK ] on the function bar.
ð The display shows a dialog with an alphanumeric input box.
u Enter a name for the set-up.
u Tap the function [ Save ] on the function bar.
ð The name is saved.
ð The dialog box closes.
6.6 Editing a method
The Pro interface can save up to 40 extraction methods.
6.6.1 Creating a new method
Navigation path
➔
u Navigate to the [ Method ] menu via the navigation path.