BÜCHI Labortechnik AG Operating the interface pro | 6
Operation Manual UniversalExtractor E-800 57/88
u Tap the action [ Sox valve open time ].
ð The display shows a dialog with the selectable values.
u Tap the value you wish to use.
u Tap the function [ OK ] on the function bar.
ð The dialog box closes.
u Tap the function [ Save ] on the function bar.
ð The Soxhlet valve open time is saved.
6.6.11 Setting the rinse time of a method (Soxhlet, Soxhlet Warm, Hot and
Twisselmann Extraction)
The duration of the rinse time is selectable from 1 to 5940 min.
Navigation path
➔
u Navigate to the [ Method ] menu via the navigation path.
u Tap the name of the method that you wish to edit.
ð The display shows the view Method.
u Tap the function [ Edit ] on the function bar.
u Tap the action [ Rinse time ].
ð The display shows a dialog box with a numeric input box.
u Enter the value in the numeric input box.
u Tap the function [ OK ] on the function bar.
ð The dialog box closes.
u Tap the function [ Save ] on the function bar.
ð The rinse time is saved.
6.6.12 Setting the rinse heating level of a method (Soxhlet, Soxhlet Warm, Hot and
Twisselmann Extraction)
The default setting for the rinse heating level is depending on the selected solvent.
Change the rinse heating level to compensate fluctuations in air pressure.
The heating steps are selectable from 0 to 20.
Navigation path
➔
u Navigate to the [ Method ] menu via the navigation path.
u Tap the name of the method that you wish to edit.
ð The display shows the view Method.
u Tap the function [ Edit ] on the function bar.
u Tap the action [ Rinse heating level ].
ð The display shows a dialog box with a numeric input box.
u Enter the value in the numeric input box.