6 | Operating the interface pro BÜCHI Labortechnik AG
60/88 Operation Manual UniversalExtractor E-800
u Tap the name of the solvent you wish to edit.
ð The display highlights the selection in green.
u Tap the function [ OK ] on the function bar.
ð The display shows the view Solvent.
u Tap the action [ Heating level ].
ð The display shows a dialog box with a numeric input box.
u Enter a heating level for the solvent.
u Tap the function [ Save ] on the function bar.
ð The new name is saved.
ð The dialog box closes.
6.7.5 Changing the boiling point of a solvent
Navigation path
u Navigate to the [ Solvent ] menu via the navigation path.
u Tap the name of the solvent you wish to edit.
ð The display highlights the selection in green.
u Tap the function [ OK ] on the function bar.
ð The display shows the view Solvent.
u Tap the action [ Boiling point ].
u Tap the function [ OK ] on the function bar.
ð The control panel shows a dialog box with a numeric input box.
u Enter the boiling point for the solvent.
u Tap the function [ OK ] on the function bar.
ð The value for the boiling point is saved.
ð The dialog box closes.
6.7.6 Changing the enthalpy of a solvent
Navigation path
u Navigate to the [ Solvent ] menu via the navigation path.
u Tap the name of the solvent you wish to edit.
ð The display highlights the selection in green.
u Tap the function [ OK ] on the function bar.
ð The display shows the view Solvent.
u Tap the action [ Enthalpy ].
u Tap the function [ OK ] on the function bar.
ð The display shows a dialog box with a numeric input box.
u Enter the enthalpy for the solvent.