6 | Operating the interface pro BÜCHI Labortechnik AG
58/88 Operation Manual UniversalExtractor E-800
u Tap the function [ OK ] on the function bar.
ð The dialog box closes.
u Tap the function [ Save ] on the function bar.
ð The rinse heating level is saved.
6.6.13 Setting Dry step
NOTICE
Risk of beaker damage or deterioration of the analyte due to incorrect
settings of the drying step.
Incorrect drying step settings can damage the analyte and the beaker.
u Use the Analyte Protection for the first drying step.
u Changing the default settings is at your own risk.
A maximum of three dry steps are possible.
Adding a dry step by tap the [ + ] button.
For the dry step the following settings are possible:
Action Option Explanation
Analyte Protection
AP
On/off Enables or disables the analyte protection.
Time Enter value Heating time for the individual step.
Heating level Enter value Heating level for the individual step.
Navigation path
u Navigate to the [ Method ] menu via the navigation path.
u Tap the name of the method that you wish to edit.
ð The display shows the view Method.
u Tap the function [ Edit ] on the function bar.
u Tap the action [ Dry step ].
u Enter the drying step configuration according to you needs.
u Tap the function [ OK ] on the function bar.
u Tap the function [ Save ] on the function bar.
ð The dry step configuration is saved.