74
18. Wait for the instrument to tare the balance.
Start Weight
SAMPLE ID: XXXX
CRUCIBLE WT: XXXX.X
Place syringe back onto
balance pan and press NEXT
Press BACK to Retare
Note: Press the BACK key to retare the balance.
19. Place the empty syringe back on the external balance pan.
20. Press the NEXT key.
READING BALANCE
21. Wait for the instrument to determine the weight of the empty syringe.
Start Weight
SAMPLE ID: XXXX
SAMPLE WT: XXX
Press NEXT key to weigh next
sample, or press SELECT to
return to run screen.
22. Press the SELECT key.
XXXXXXXX
RAMP XX C
XX:XX:XX Set Point
DWELL XXX C
XX:XX:XX
Status: Paused
WARNING
To prevent the possibility of serious burns, ensure that insulated
gloves and protective gear as outlined in the user’s safety program
are worn and metal tongs are used to insert and remove sample con-
tainers from the ashing furnace.
23. Open the instrument door. Remove the furnace door and place it in the furnace
door holder.