8 ChroMate™ Operator’s Manual Doc. 4300 Rev. E
2. TECHNICAL SPECIFICATIONS
The plate moves toward the front and back of the instrument.
When taking an optical reading, the reaction plate automatically positions itself under the
8-channel optical system. A filter wheel with four or six filters rotates below the plate.
Depending on the setup, reports may be displayed or printed to create permanent lab
records and physician reports.
Photometer:
Optical Measurement Range: -0.20 to 4.0 Absorbance Units (A)
Photometric Accuracy: +/- 1% or better (NIST)
Stability: Drift of no more than 0.005A in 8 hours
Warm up: 45 seconds
Light Source: Halogen, lamp saver feature
Wavelengths: Standard four filter: 405, 450, 492, and 630nm
6-filter: 405, 450, 492, 545, 600, and 630nm
Filter options from 405nm to 750nm
Filter Type: IAD hardcoat, 10nm half bandpass
Vessel: Standard 96 well microtiter plates or strip trays
Electronic:
Power Requirements: 100 - 240VAC, 50-60Hz 90W maximum.
All power cords must be approved for the
country of use
PC Connection: USB Port
Microprocessor: eZ80
Speed: Reads absorbance of 96 wells in
approximately 8 seconds
Plate Transport: Stepper motor
Certifications: Pending