IOLITE®
TECHNICAL REFERENCE MANUAL
4.4.2.1. IOLITEi-8xTH-HS: Specifications
Isolated universal thermocouple and voltage
Simultaneous 10, 20, 40, 50, 80, 100, 200, 500, 1k, 2k, 5k, 10k S/sec
±0.02 % of reading ±50 μV
Typical Noise floor @ 10/100/1k/10k S/sec
-130 dB / -120 dB / -110 dB / -104 dB
-120 dB / -110 dB / -100 dB / -90 dB
Typical 0.05 μV/K (max. 0.2 μV/K)
Typical 0.01 μV/K (max. 0.1 μV/K)
Typical 2 ppm/K (max. 10 ppm/K)
TC Types: K, J, T, R, S, N, E, C, B
±0.02 % of reading ±0.5 °C ±7 μV
0.002 °C RMS@Type K @ 10 S/sec
0.006 °C RMS@Type K @ 100 S/sec
0.02 °C RMS@Type K @ 1k S/sec
0.05 °C RMS@Type K @ 10k S/sec
Mini Thermocouple connector (copper)
1000 Vpeak channel to ground & channel to channel
Typical 3.2 W (max. 3.4 W)