Figure 41 – Added sample sedimentation time
Note: if samples are not loaded within 30 seconds of opening the cover, as a safety
feature, the entire cycle or working session will be aborted. The bottom of the display
will show “Cycle #X completed – Ready” and the instrument will not deduct runs from
the counter.
Important: The ‘ADD’ button is disabled (gray and not selectable) if the remaining
time for reading the sample on board is 2 minutes. Is not possible to add samples
with random access but the user has to wait.
At the end of the cycle, the instrument allows the user to recover the barcodes which
it was not able to scan previously (three short audible alarm warnings) with the same
steps described above.
Figure 42 - Barcode insertion at the end of the cycle
A cycle has ended when the message 'Cycle xx complete - Ready' can be read at the
bottom of the display. The instrument is ready for a new work session.