Instructions for use D-Vapor/D-Vapor 3000 47
Preparation
The bottle valve is now open. Hold the bottle firmly
in this position. Anesthetic agent will flow into the
vaporizer. The contents of the bottle will show
bubbles. Watch the filling level in the sight glass. Fill
the vaporizer no higher than the maximum mark.
If the Vaporizer is filled when it is warmed up, the
contents of the tank will also bubble. Briefly
interrupt the filling to check the fill level.
If the vaporizer is filled beyond the maximum mark,
excess anesthetic agent may spray from the filling
port when the bottle is removed.
Ending the filling procedure
1 Slowly turn the anesthetic agent bottle (B)
counterclockwise.
The spring in the integrated closure (C) ensures
that the bottle valve closes and prevents loss of
anesthetic agent. The filling procedure is
completed.
2 Continue to slowly turn the anesthetic agent
bottle and withdraw it from the filling port.
3 Check the filling level in the sight glass. The
vaporizer must be hanging or standing vertically
during the check. The filling level must be
visible and must not be above the maximum
mark.
4 Place the screw cap (A) in the filling port and
screw it all the way in.
5 Close the anesthetic agent bottle with the cap.
CAUTION
Health hazard
A bottle which was dropped may release signifi-
cant quantities of anesthetic agent, if broken.
Do not use a broken bottle and remove it from the
room.
076
0
B
C
077078
0
Vol.%
B
0
Vol.%
A