Problem solving
80
Instructions for use D-Vapor/D-Vapor 3000
Note Yellow Battery LED is lit. Battery is not ready for the
specified battery operation.
Connect vaporizer to the
mains power supply without
dosing (control dial to posi-
tion 0 or T).
If the yellow Battery LED is
still lit after 1 hour, the bat-
tery is defective.
Have the battery replaced
by service personnel,
see page 38.
Repair by specialized ser-
vice personnel.
Note Yellow Fill Up LED is lit. The level is below the refill
mark.
A full bottle of desflurane
anesthetic agent can be
added, see "Filling the Va-
porizer", page 42.