137
Technical Publication Document NO: SYSTEM-007-ENG (Ver.1.1)
HF-525Plus Installation, Operation and Service Manual
6.1.4 When to do calibration
After product installation,
including relocation
Calibration for X-ray
exposure parameters
Periodic testing for
maintenance
(Every six months or as
required)
Calibration for X-ray
exposure parameters
Replacement or repair of
critical components
Calibration for X-ray
exposure parameters
Refer to the chapter 4 and 6
6.1.5 Critical Components that could affects the product performance
Replacement or repair of critical components requires a calibration for X-ray exposure parameters.
Critical components are the followings:
● X-Ray tube assembly
● MiO Board
● CnD Board
● FnD Board
● Transformers of high voltage
● High voltage cables