Technical Publication Document NO: SYSTEM-007-ENG (Ver.1.1)
HF-525Plus Installation, Operation and Service Manual
6.1.4 When to do calibration
When to do
What to do
Remarks
After product installation,
including relocation
Calibration for X-ray
exposure parameters
Refer to the chapter 4
Periodic testing for
maintenance
(Every six months or as
required)
Calibration for X-ray
exposure parameters
Refer to the chapter 4
Replacement or repair of
critical components
Calibration for X-ray
exposure parameters
Refer to the chapter 4 and 6
6.1.5 Critical Components that could affects the product performance
Replacement or repair of critical components requires a calibration for X-ray exposure parameters.
Critical components are the followings:
X-Ray tube assembly
MiO Board
CnD Board
FnD Board
Transformers of high voltage
High voltage cables
6.1.6 Installation of “USB to RS232 Converter Driver”
Specifications
Identification: USB to RS232 Converter
Confirmation of PC specifications
√ System: Pentium MMX 200MHz or higher
√ Operating System: Windows XP Service Pack2 or higher
√ USB Interface: USB 2.0 or higher (USB 2.0 recommended)
The software for RS232 Driver is available on the website and the installation method refers to a
manual which is accompanied the RS232 Converter.
Installation
Turn on PC.
Click “My Computer” icon from the wallpaper on PC to execute “Properties”.
Click “Properties”.