222
Technical Publication Document NO: SYSTEM-007-ENG (Ver.1.1)
HF-525Plus Installation, Operation and Service Manual
Appendix 1 EMC – Manufacturer’s Declaration
HF-525 PLUS is manufactured to be used in the electromagnetic environment designated in the below table.
The user of HF-525 PLUS should check whether the equipment is used in such an environment.
Electromagnetic Environment - Guide
HF-525 PLUS uses RF energy only for internal
functions. As a result, RF emission is very low and
does not interfere with surrounding electronic
devices.
HF-525 PLUS is adequate for use in all facilities, and
easily utilizes the low voltage power supply provided
to residential facilities and buildings.
Harmonic wave emission
IEC 61000-3-2
Voltage variations / Flicker
emission IEC6100-3-3
Electromagnetic Susceptibility
HF-525 PLUS is manufactured to be used in the electromagnetic environment designated in the below table.
The user of HF-525 PLUS should check whether the equipment is used in such an environment.
Electromagnetic Environment -
Guide
Electrostatic
Discharge (ESD)
IEC 61000-4-2
± 6 kV Contact
± 8 kV Stand-by
The flooring should be wood,
concrete, or ceramic tile. If the floor
is covered with synthetic materials,
the relative humidity should be over
30%.
Electrical Transient
/ Burst
IEC 61000-4-4
± 2 kV (for power supply
line)
± 1 kV (for I/O line)
The quality of the main power
supply should be equivalent to that
of commercial or hospital settings.
± 1 kV (line-line)
± 2 kV (line-earth)
The quality of the main power
supply should be equivalent to that
of commercial or hospital settings.