EasyManua.ls Logo

Ecoray HF-525Plus - Start Calibration

Ecoray HF-525Plus
233 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
155
Technical Publication Document NO: SYSTEM-007-ENG (Ver.1.1)
HF-525Plus Installation, Operation and Service Manual
6.6.6 Start calibration
Install equipments for calibration. (Refer 3 of this chapter)
Execute “Calibration window”.
6.6.7 Selection of X-ray tube and small focal
Select a suitable X-ray tube and small focal.
Check the selected of focal (see No 2 of the above fig).
Click the No 5 related to No 4 for min. kV and No 3 for min. mA (see No 3, 4, 5 of the above fig).
When click the No 5, the colume changes to yellow colour. (see No 5 of the above fig).
Check the selected data (see No 6 of the above fig).
Execute “X-ray exposure”.
Note: Under pushing “Ctrl” on keyboard for “Ready”, pushing “X” on keyboard for “X-ray exposure”
Check the detected kV and mA (see No 6 of the above fig).
On the “ADJ”, change data (see No 6 of the above fig).
Click “Enter” on keyboard.
Execute “X-ray exposure”.
When the detected kV and mA are not incorrect, execute “X-ray exposure” again and change data.
Click “Enter” on keyboard.

Table of Contents

Related product manuals