EasyManua.ls Logo

Ecoray HF-525Plus - Page 154

Ecoray HF-525Plus
233 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
154
Technical Publication Document NO: SYSTEM-007-ENG (Ver.1.1)
HF-525Plus Installation, Operation and Service Manual
Click “Calibration On/Off”(see No 1of the above fig).
The “System State” indicates “Now Data Access” (see No 5 of the above fig).
When it indicates“Now Data Access”, the calibration data of MiO board move to a calibration program
of PC automatically (see No 5of the above fig).
Note: MiO board means a “Main Input & Output Board” of generator
When this calibration data has been moved, data table indicates the calibration data (see No 7of the
above fig).
Select a suitable X-ray tube and focal spot (see No 2of the above fig).
Check the selected X-ray tube and focal spot (see No 6of the above fig).
The ““Data Clear” and “Current Focal Load” are only used in the manufacturing process (see No 3 of
the above fig).
Click “Store” for saving input data (see No 4 of the above fig).

Table of Contents

Related product manuals