184
Technical Publication Document NO: SYSTEM-007-ENG (Ver.1.1)
HF-525Plus Installation, Operation and Service Manual
√ Check whether the x-ray system is activated after the console "Power On".
√ Check that the console's Standby Lamp is on. (Normal: yellow)
√ Check if the Waiting Lamp turns off when the console's power is "ON". (Normal: Off)
√ If the power of the console is "ON", check if the Standby Lamp lights up in yellow and green and
lights in green. (Normal: green light)