EasyManua.ls Logo

Ecoray HF-525Plus - Page 196

Ecoray HF-525Plus
233 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
196
Technical Publication Document NO: SYSTEM-007-ENG (Ver.1.1)
HF-525Plus Installation, Operation and Service Manual
12.3.13 Error type 13: Displayed E r r FP at kVp, mA, Time section
12.3.14 Error type 14: Displayed E r r FILA at kVp, mA, Time section
12.3.15 Error type 15: Displayed E r r 232 at kVp, mA, Time section
12.3.16 Error type 16: Displayed E r r AEC at kVp, mA, Time section
12.3.17 Error type 17: Displayed E r r SEC at kVp, mA, Time section
12.3.18 Error type 18: Displayed E r r Pro----O at kVp, mA, Time section
12.3.19 Error type 19: Displayed E r r Pro--I-- at kVp, mA, Time section
Circuit
error
Stator
Circuit
error
Circuit
error
HV-DRV
IGBT-OCP
DRV-PER
Filament
- OLP
nication
error
Circuit
error
time error
Normal
To be
checked
Normal
Normal
Normal
Normal
Normal
Charge
Circuit
error
Tube
Stator
Circuit
error
Filament
Circuit
error
Refer to XR7-FnD Circuit
Commu
nication
error
AEC
Circuit
error
Exposure
time error
HV-DRV
IGBT-OCP
DRV-PER
Filament
- OLP
Normal
Normal
To be
checked
To be checked
(Any one of 4 is possibly faulty)
Normal
Normal
Normal
Charge
Circuit
error
Tube
Stator
Circuit
error
Filament
Circuit
error
Refer to XR7-FnD Circuit
Commu
nication
error
AEC
Circuit
error
Exposure
time error
HV-DRV
IGBT-OCP
DRV-PER
Filament
- OLP
Normal
Normal
To be
checked
Normal
Normal
Normal
Normal
Charge
Circuit
error
Tube
Stator
Circuit
error
Filament
Circuit
error
Refer to XR7-FnD Circuit
Commu
nication
error
AEC
Circuit
error
Exposure
time error
HV-DRV
IGBT-OCP
DRV-PER
Filament
- OLP
Normal
Normal
Normal
Normal
To be
checked
Normal
Normal
Charge
Circuit
error
Tube
Stator
Circuit
error
Filament
Circuit
error
Refer to XR7-FnD Circuit
Commu
nication
error
AEC
Circuit
error
Exposure
time error
HV-DRV
IGBT-OCP
DRV-PER
Filament
- OLP
Normal
Normal
Normal
Normal
Normal
To be
checked
Normal
Charge
Circuit
error
Tube
Stator
Circuit
error
Filament
Circuit
error
Refer to XR7-FnD Circuit
Commu
nication
error
AEC
Circuit
error
Exposure
time error
HV-DRV
IGBT-OCP
DRV-PER
Filament
- OLP
Normal
Normal
Normal
Normal
Normal
Normal
To be
checked
Charge
Circuit
error
Tube
Stator
Circuit
error
Filament
Circuit
error
Refer to XR7-FnD Circuit
Commu
nication
error
AEC
Circuit
error
Exposure
time error
HV-DRV
IGBT-OCP
DRV-PER
Filament
- OLP
Normal
Normal
Normal
To be
checked
Normal
Normal
Normal
Normal
Normal
Normal
Charge
Tube
Filament
Refer to XR7-FnD Circuit
Commu
AEC
Exposure

Table of Contents