48
Technical Publication Document NO: SYSTEM-007-ENG (Ver.1.1)
HF-525Plus Installation, Operation and Service Manual
After Fastening HV Cable Plug nut and Using Wrench Volt, Completely Fastening HV Cable Plug.
WARNING
The terminal pins of the high voltage cable are extremely delicate and easily damaged. And then,
handle them with very carefully.
WARNING
Always Maintaining Clean for HV Cable Plug and Receptacle socket, and have a caution for sticking
with Dirt
.
WARNING
After Inserting HV Cable Plug to Receptacle socket, and completely fastening Ring nut and Wrench
volt, if by Faulty Connection of Socket, It Can Cause HV Leakage for Broken Device.
5.9.5.7 X-Ray Console Cable Connection