Acclarix AX8/Acclarix AX7 Diagnostic Ultrasound System User Manual Specifications
- 170 -
Appendix A Specifications
A.1 Electrical Safety Classifications
According to the type of protection
against electric shock
Internally powered equipment,
Class I equipment
According to the degree of protection
against electric shock
Type BF
Type CF(ECG module)
According to the degree of protection
against harmful ingress of liquid
Whole device: IPX0
Console (including touch screen, track pad and hard keys):
IPX1
Transducer (not including the transducer connector): IPX7;
Footswitch: IP68
According to the degree of safety of
application in the presence of a
flammable gas
Equipment not suitable for use in the presence of a
flammable gas
According to the mode of operation
According to the grade of EMC
CISPR 11 Group 1, Class A
EN 60601-1:2006/A1:2013
idt IEC 60601-1: 2005/A1:2012
EN 60601-1-2: 2015
idt IEC 60601-1-2: 2014
EN 60601-2-37:2008/A1:2015
idt IEC 60601-2-37:2007/A1:2015
IEC 62359: 2017
A.2 Power Supply
A.3 Battery
About5.5 hours with system power-on; About 2.5 hours with
system power-off