Acclarix AX8/Acclarix AX7 Diagnostic Ultrasound System User Manual Specifications
- 173 -
Resolution: 1 bpm
Accuracy: ±1% or 1 bpm, whichever is greater.
QRS Detection
Range: 40 ms~120 ms
Amplitude:0.5 mV~5 mV
Patient Leakage
Current
< 10 μA
Supply Voltage
USB-supply (5V, 300mA)
A.7 Operating, Storage and Transportation Environment
A.7.1 Operating Environment
Temperature
0 °C ~ +40 °C(+32 °F~+104 °F)
Relative humidity range
15% RH ~ 95% RH(non-condensing)
Atmospheric pressure range
86kPa ~ 106kPa
A.7.2 Storage and Transportation Environment
Temperature
-20 °C ~ +55 °C(-4 °F~+131 °F)
Relative humidity range
15% RH ~ 95% RH(non-condensing)
Atmospheric pressure range
70kPa ~ 106kPa
A.8 Transducer Specifications
No.
Transducer
Center Frequency
No.
Transducer
Center Frequency
1
L17-7HQ
12.0MHz
2
E8-4Q
6.2MHz
3
C5-2MQ
3.5MHz
4
C5-1Q
3.25MHz
5
C5-2Q
3.5MHz
6
L12-5Q
8.0MHz
7
L17-7SQ
12.0MHz
8
MC8-4Q
6.5MHz
9
P5-1Q
2.8MHz
10
P7-3Q
5.0MHz
11
MC9-3TQ
6.4MHz
12
C7-2XQ
4.2MHz
13
E10-3BQ
6.5MHz
14
E10-3HQ
6.5MHz
15
C5-2XQ
3.5MHz
16
L10-4Q
6.8MHz
17
P5-1XQ
2.8MHz

Table of Contents