Acclarix AX8/Acclarix AX7 Diagnostic Ultrasound System User Manual Getting Started
- 22 -
3.3 System Preparation
3.3.1. Battery Use
The system comes with one lithium-ion battery. When it is fully charged it can run the system for
approximately 60 minutes, depending on use. It is automatically charged when the system is plugged
in.
CAUTION
1. If the system will remain unused for more than one week, charge the battery to at least 75%
capacity, take the battery out and store the system and battery separately.
2. During long term storage, the battery should be charged at least once every 3 months to ensure
battery capacity is more than 75%.
3. Only use Edan supplied battery.
To install the battery:
1. Turn off and unplug the system.
2. Close the monitor, turn the system upside down and rest it on a flat stable surface.
3. Put the battery gently to battery compartment, and the button will lock the battery automatically.
Figure 3-6 Installation of the Battery
WARNING
1. When the battery capacity is ≤20%, the battery status icon turns red.
2. When the battery capacity is 10%, the system displays a prompt Low Battery. Please plug in
the adapter to ensure uninterrupted use."
To remove a battery:
1. Turn off and unplug the system.
2. Close the monitor, turn the system upside down and rest it on a flat stable surface.
3. Slide the battery button to unlock the battery.
4. Take the battery out.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN Acclarix AX7 and is the answer not in the manual?

EDAN Acclarix AX7 Specifications

General IconGeneral
BrandEDAN
ModelAcclarix AX7
CategoryDiagnostic Equipment
LanguageEnglish

Related product manuals