M3 Vital Signs Monitor User Manual Specifications
- 97 -
TH Module:
Measuring range
34 C~ 42.2 C
Working temperature
10 C~ 40 C
Adjustable range of alarm limits
35.5 C~ 42 C
Resolution
0.1 C
Response time
1 s
Measuring Mode
Adjusted Mode
Clinical Accuracy
0.2 C (0.4 ° F) (35.5 C ~42 C) (95° F
~107.6°F)
0.3 C (0.5°F) (out of the range mentioned
above)
Laboratory Accuracy
0.2 C
F3000 Module:
Measuring range
30 C~43 C
Prediction measurement range
35 C~43 C
Cold mode prediction measurement
range
33 C~43 C
Working temperature
10 C ~ 40 C
Sensor type
Oral /axillary /rectal
Adjustable range of alarm limits
33 C~43 C
Resolution
0.1 C
Accuracy
Monitoring Mode and Predictive Mode: 0.1 C
Quick Predictive Mode: 0.3 C
Typical measurement time (after
insertion into measurement site)
Oral (Quick Predictive Mode): (3~5) s (non-fever
temps); (8~10) s (fever temps)
Oral (Predictive Mode): (6~10) s
Axillary: (8~12) s
Rectal: (10~14) s
Monitoring Mode (all sites): (60~120) s