M3 Vital Signs Monitor User Manual System Menu
- 30 -
alarms will be disabled; medium and low technical alarms will only be indicated by
turns in the information area. In the monitoring mode, audio and visual prompts for all
alarms will be effective and alarms will be indicated by turns in the information area.
3 In the spot check mode, medium and low technical alarms cannot disable the
settings of silencing the alarm. Only when a new physiological alarm or a high
technical alarm occurs can the monitor automatically exit the alarm silenced status.
In the monitoring mode, if a new alarm of any type occurs, the monitor will
automatically exit the alarm silenced status.
4 In the spot check mode, no trend graph will be shown.
WAVE FORM: Set displayed waveforms to UNFILLED or FILLED.
FACE SELECT: Set NIBP or SpO
2
as the main displayed parameter onscreen (FACE
SELECT is only available for the monitor with the configured modules NIBP+SpO
2
).
RECORDER SPEED: Set recorder speed to 12.5 mm/s or 25 mm/s.
EXIT: Return to the previous menu.
4.4 Selection
For the monitor outfitted with SpO
2
, NIBP and TEMP modules, you may select SELECTION in
SYSTEM MENU to access this submenu, in which seven selections are available: NIBP
TREND TAB, SpO
2
TREND TAB, TEMP TREND TAB, ALL PARAMETERS, ALARM
LIST, TREND GRAPH and TREND TABLE. Only one item can be selected to display
information on the interface.
NIBP TREND TAB: to display NIBP trend table;
SpO
2
TREND TAB: to display SpO
2
trend table;
TEMP TREND TAB: to display TEMP trend table;
ALL PARAMETERS: to display all parameters in the area;
ALARM LIST: to display alarm list;
TREND GRAPH: to display the trend graph;
TREND TABLE: to display the trend table.
You can shift the data list to waveform display by pressing the TREND/WAVEFORM button
on front panel.
For Single display mode, the Selection menus are different.
4.5 Deleting Data
If you press the button when a trend list or an alarm list is displayed on the screen, it
will display below selections:
DELETE ID: Entirely delete the trend and alarm data of the current monitored patient.
DELETE ALL DATA: Entirely delete the trend and alarm data of all the monitored patients.