EasyManuals Logo

EDAN M3 User Manual

EDAN M3
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #66 background imageLoading...
Page #66 background image
M3 Vital Signs Monitor User Manual SpO
2
Monitoring (Optional)
- 58 -
8.8.2 SatSeconds Safety Net
The SatSeconds “Safety Net” is for patients whose saturation makes frequent excursions below or
above the SpO
2
limit but does not remain in violation long enough for the SatSeconds limit to be
reached. If three or more SpO
2
alarm limit violations occur within a 60-second period, an alarm
will be triggered even if the SatSeconds limit has not been reached.
8.8.3 Setting SatSeconds Duration
You can set SatSeconds to Off or to the duration among 10, 25, 50 and 100. To configure the
SatSeconds settings, enter the SpO
2
Setup menu and select the desired SatSeconds setting from
the SatSeconds list.
8.9 Alarm
8.9.1 Adjustable Range of Alarm Limits
SpO
2
/ PR adjustable range of alarm limits:
Adjustable Range
SpO
2
0~100
PR
30~300
Default SpO
2
alarm limits:
Max. Upper Limit
Min. Lower Limit
Step
ADU
100
90
1
PED
100
90
1
NEO
95
88
1
Default PR alarm limits:
Max. Upper Limit
Min. Lower Limit
Step
ADU
120
50
1
PED
160
75
1
NEO
200
100
1
8.9.2 Alarm Information and Prompts
Physiological alarm:
Message
Cause
Alarm Level
SpO
2
HIGH
SpO
2
measuring value is above upper alarm limit.
Medium
SpO
2
LOW
SpO
2
measuring value is below lower alarm limit.
Medium
PR HIGH
PR measuring value is above upper alarm limit.
Medium
PR LOW
PR measuring value is below lower alarm limit.
Medium

Table of Contents

Other manuals for EDAN M3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN M3 and is the answer not in the manual?

EDAN M3 Specifications

General IconGeneral
BrandEDAN
ModelM3
CategoryMonitor
LanguageEnglish

Related product manuals