EasyManua.ls Logo

EDAN M3 - TEMP Monitoring with TH Module; Introduction

EDAN M3
116 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
M3 Vital Signs Monitor User Manual TEMP Monitoring (Optional)
- 73 -
Prompt:
Message
Cause
Warm-up over
The monitor prompts it after taking the sensor out of the bracket and
warm-up is over.
Measure over
After the Predict measuring is over, the data and message display on
the interface.
Measure time out
No measuring result after the module entering Predict state for 30 s.
Probe temp too high
The original temperature of sensor >+33 C and +40 C.
Temp SENSOR OFF
The probe is disconnected from the patient.
10.2 TEMP Monitoring with TH Module
10.2.1 Introduction
M3 with the TH module (Infrared Ear Temperature Module) takes a temperature in the ear.
Diagram of the Infrared Ear Thermometer
WARNING
1 The infrared ear thermometer is not intended for neonatal patients.
2 Only use the disposable probe covers supplied or recommended by EDAN. Use of
other manufacturers probe covers, reuse of disposable probe covers or absence of
probe covers may produce temperature measurement errors and/or inaccuracies.
Batter Cover
Pin Hole

Table of Contents

Other manuals for EDAN M3

Related product manuals