WARNING
Prolonged non-invasive blood pressure measurements in automatic mode may be
associated with purpura, ischemic and neuropathy in the limb wearing the cuff. When
monitoring a patient, examine the extremities of the limb frequently for normal color,
warmth and sensitivity. If any abnormality is observed, stop the blood pressure
measurements.
NOTE:
1 The width of the cuff is either approximately 40% of the limb circumference or 2/3 of
the upper arm length. The inflatable part of the cuff should be long enough to encircle
80%-100% of the limb. The wrong size of cuff can cause erroneous readings. If the
cuff size is in question, use another cuff with suitable size to avoid errors.
2 If an NIBP measurement is suspect, repeat the measurement. If you are still
uncertain about the reading, use another method to measure the blood pressure.
3 Please make sure the cuff is well connected. A leak of air may cause measurement
error.
4 Please select the cuff with the suitable size. An unsuitable cuff may cause incorrect
measurements.
5 Avoid incursion of liquid into the cuff. If this happens, please desiccate the cuff
completely.
9.4.1 Operation Prompts
1. To start auto measurement:
Access NIBP SETUP menu and pick the INTERVAL item, in which the user may set up the
time interval for auto measurement. After that, press the NIBP START/STOP on the front
panel to start the auto measurement according to the selected time interval.
2. To stop auto measuring:
During auto measuring, press the NIBP START/STOP on the front panel at any time to stop
the auto measurement in process.
WARNING
If you repeatedly use AUTO measuring in a short term, it may lead to inaccurate readings
or endanger patient’s life.
3. To start a manual measuring:
Access NIBP SETUP menu and pick the INTERVAL item. Select the MANUAL
selection. Then press the NIBP START/STOP on the front panel to start a manual
measurement.