M3 Vital Signs Monitor User Manual EMC Information
- 99 -
Appendix B EMC Information - Guidance and
Manufacture’s Declaration
B.1 Electromagnetic Emissions
Guidance and manufacture’s declaration – electromagnetic emission
M3 is intended for use in the electromagnetic environment specified below. The customer or
the user of M3 should assure that they are used in such an environment.
Electromagnetic environment – guidance
M3 uses RF energy only for their internal
function. Therefore, their RF emissions are very
low and are not likely to cause any interference
in nearby electronic equipment.
M3 is suitable for use in all establishments, other
than domestic establishments and those directly
connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
NOTE:
The EMISSIONS characteristics of M3 make it suitable for use in industrial areas and
hospitals (CISPR 11 class A). If it is used in a residential environment (for which CISPR
11 class B is normally required) M3 might not offer adequate protection to
radio-frequency communication services. The user might need to take mitigation
measures, such as relocating or re-orienting the equipment.