M3 Vital Signs Monitor User Manual Specifications
- 91 -
Appendix A Specifications
A.1 Classification
ClassⅠequipment and internal powered equipment
SpO
2
, NIBP: BF
TEMP: CF (T2 module)
BF (TH module, F3000 module)
IPX1
With T2, TH or F3000 TEMP module: Ordinary
equipment (Sealed equipment without liquid proof)
Continuous operation equipment
IEC 60601-1: 2005+A1:2012; IEC 60601-1-2:2014;
EN 60601-1: 2006+A1: 2013; EN 60601-1-2: 2015;
IEC 80601-2-30: 2009+A1: 2013; ISO 80601-2-61:
2011; ISO 80601-2-56: 2009; IEC 60601-2-49:2011
A.2 Specifications
NOTE:
The performance of the equipment with ☆ mark is determined to be essential
performance.
A.2.1 Size and Weight
(174±2) mm (W)× (235±2) mm (H)×(189±2) mm(D)
≤3.5 kg (not including battery)
A.2.2 Function Configuration
A.2.3 Environment
The monitor may not meet the performance specifications given here if stored or used outside the
specified temperature and humidity ranges.