M3 Vital Signs Monitor User Manual System Menu
- 32 -
of EDAN.
SERVER PORT: Set server port.
PRES UNIT: Set the pressure unit to mmHg or kPa.
TEMP UNIT: Set the temperature unit to C or℉.
HOT KEY: Set the hot key to PATIENT or PRINT.
COLOR SELECT: Set the color of displayed waveforms. 16 kinds of colors can be selected.
Click on DEFAULT to return to the default configuration.
OTHER SETUP
SpO
2
SETUP:
SpO
2
ALARM LEV
You can configure the alarm level for SpO
2
SENSOR OFF to HIGH or LOW.
SENSITIVITY
The SpO
2
reading is the average of data collected within a specific time. You can set
Sensitivity to HIGH, MED or LOW via the menu. The higher the sensitivity is, the quicker
the monitor responds to the changes in the patient’s oxygen saturation level. Contrarily, the
lower the sensitivity is, the slower the monitor responds to the changes in the patient’s
oxygen saturation level, but the measurement accuracy will be improved. When a critical
patient is monitored, selecting high sensitivity will help to understand the patient’s state.
PI
PI is a numeric value indicating perfusion level. It reflects the perfusion level at the monitoring
site. You can set it ON or OFF.
NIBP SETUP:
AVG INTERVAL: Set the average NIBP measurement interval to 1MIN, 2MIN,
3MIN, 4MIN or 5MIN.
AVG TIMES: Set the average NIBP measurement times to 3 or 5.
RESET: Select it to reset the NIBP module.
- Restore measurement status.
- Pick this item to restore initial settings of the pressure pump.
- When the pressure pump does not work properly and the system fails to give message for
the problem, pick this item to activate self-test procedure, thus restore the system from
abnormal performance.
LEAK TEST
This item is used for an air leakage test. Select this item to start the air leakage test. Then the
item will change into STOP LEAK TEST. Select it again, and the system will stop the air
leakage test.