SE-1201 Electrocardiograph User Manual System Setup
- 102 -
10.10 File Setup
Select File on the System Setup screen, and press Enter to open the File Setup window.
Figure 10-11 File Setup Window
Item Description
Auto Save
Choose from: Off, To ECG or To External Memory
Select Off, ECG data will not be saved.
Select ECG, ECG data in the auto, rhythm or off mode will be saved in
the ECG automatically.
Select To External Memory, ECG data in the auto or rhythm mode will
be automatically saved to the directory of
ECGDATA\ECG-X\Store\Examination Date of the U disk or SD card
after an ECG report is printed out. In the off mode, 10s ECG data sampled
after pressing the PRINT/STOP key will be automatically saved to the
directory of ECGDATA\ECG-X\Store\Examination Date of the U disk or
SD card.
NOTE:
1. Please insert the U disk or SD card recommended by the
manufacturer. Please set the format to FAT or FAT32 when
formatting the U disk or SD card.
2. X in the directory of ECGDATA\ECG-X\Store\Examination
Date can be set in the Device No. textbox in the Transmission
Setup window.
File Format
Choose from: DAT, SCP, FDA-XML or PDF
To select SCP\FDA-XML, you should first activate the SCP/FDA-XML
function on the Advanced Setup screen. For details on activating the
SCP/FDA-XML function, please contact the manufacturer or the local
distributor.