SE-1201 Electrocardiograph User Manual Printing ECG Reports
- 50 -
2. When the main screen is displayed, press the MODE key to select the rhythm mode. Press
F3 to select a paper speed. Press F4 to switch the gain. Press F5 to set the EMG filter or the
Lowpass filter.
3. Press the PRINT/STOP key to begin sampling, the sampling time will be displayed on the
main screen. When the sampling time reaches 60s in the Single Lead style or 20s in the
Three Leads style, it begins to print an ECG report.
4. It will stop automatically after printing a complete report of rhythm-lead ECG waveforms. Or
press the PRINT/STOP key again to stop printing the ECG report.
6.1.4 R-R Analysis Mode
Operation Method:
1. To set Mode Options or Rhythm Lead1
1) Select R-R from the Mode Option list in the Work Mode Setup window.
2) Select a lead from the Rhythm Lead1 list in the Lead Setup window.
3) Press Enter to confirm.