SE-1201 Electrocardiograph User Manual Entering Patient Information
- 44 -
First Name Within 30 ASCII characters
Last Name Within 30 ASCII characters
Age
Age Unit: Years, Months, Weeks or Days
Gender Patient Gender (Male/Female/Empty)
Pacemaker
If you select Pacemaker in the Patient Information Setup window,
Pacemaker appears in the Patient Information window. If
Pacemaker is set to Yes, the pacemaker signals are easy to be detected.
If Pacemaker is set to No, the pacemaker signals are not easy to be
detected.
BP Patient Systolic Blood Pressure/Diastolic Blood Pressure
Race
Patient Race (unknown/ Oriental/ Caucasian/ Black/ Indian/ Mongolian/
Hispanic/ Asian/ Pacific/ Chinese/ Malay/ other)
NOTE:
1) In the auto, rhythm or off mode, when ID is set to Manual and ID Hint is set to On, if
you do not input the patient ID before pressing the PRINT/STOP key, a hint will pop
up to remind you to input the patient ID.
2) The total number of supported characters may be fewer if either special Latin
characters or Chinese characters are entered.
3) If you select D.O.B in the Patient Information Setup window, the D.O.B textbox
appears and the Age textbox becomes unavailable in the Patient Information
window, you can enter the birthday of the patient, and the system will calculate the
patient age automatically.
4) If you select Age Group in the Patient Information Setup window, the Age Group
textbox appears in the Patient Information window.
5.2 Entering Patient Information by Using a Bar Code Reader
(Optional)
Operation procedures are as follows:
1. Configure the bar code
For more detailed information about configuring the bar code, please contact the
manufacturer or the local distributor.
2. Connect the bar code reader to USB socket 2 on the right panel of the electrocardiograph.