Hint information and the corresponding causes provided by the electrocardiograph are listed in
Table 11-1.
Table 11-1 Hint Information and Causes
Hint Information Causes
Lead off
Electrodes fall off the patient or the patient cable falls off the unit,
or a high polarization voltage occurs.
Battery Weak The battery is weak.
No Paper Recorder paper runs out or is not loaded.
Paper Error
When Paper Marker is set to Yes , the electrocardiograph
advances the recorder paper to the next black marker. If it
advances the paper and can not find the next black marker, the hint
Paper Error is displayed.
Testing The ECG data is being sampled periodically.
Sampling/Analyzing/
Recording
ECG signals are being sampled / analyzed / recorded.
Learning
The self-study process of arrhythmia arithmetic in the Trigger
Sample mode
Detecting
The examining process of arrhythmia data in the Trigger Sample
mode
Transmitting
ECG data is being transmitted from the electrocardiograph to the
PC through the net in the auto, rhythm or off mode.
Transmit Fail
ECG data fails to be transmitted from the electrocardiograph to the
PC through the net in the auto, rhythm or off mode.
Memory Full
The amount of files on the File Manager screen of SE-1201
reaches 200.
Module Error There is something wrong with the signal sample module.
DEMO The system is in the demonstration mode.
Overload The direct current offset voltage on an electrode is too high.
U Disk / SD Card /
USB Printer / USB
Scanner
A U disk, an SD card, a USB printer or a bar code reader is