SE-1201 Electrocardiograph User Manual System Setup
- 106 -
Item Description
External Output
The external output socket is equipped in the electrocardiograph, through
which the electrocardiograph can send signals to the external equipment.
Choose from: Off, Standard or Triggered
If External Input is set to On, and External Output is set to Standard
or Triggered, the electrocardiograph sends the signals which it receives.
If External Input is set to Off, and External Output is set to Standard,
the electrocardiograph sends ECG signals of rhythm lead 1.
If External Input is set to Off, and External Output is set to Triggered,
the electrocardiograph sends pulses with the height of 5V and the width of
40ms, based on the data of rhythm lead 1.