SE-601 Series Electrocardiograph User Manual
- 96 -
Chapter 11 Hint Information
Hint information and the corresponding causes provided by the electrocardiograph are listed in
Table 11-1.
Table 11-1 Hint Information and Causes
Hint Information Causes
Lead off Electrodes fall off the patient or the patient cable fall off the unit.
Battery Weak The built-in battery is weak.
No Paper Recorder paper runs out or is not loaded or the casing is not firmly
closed.
Paper Error In the AUTO or RHYT mode, when Paper Marker is set to Style 1
or Style 2, the recorder can not find the next paper marker at the end
of the printing course.
Sampling / Analyzing /
Recording
ECG signals are being sampled / analyzed / recorded.
Learning The self-study process of arrhythmia arithmetic in the Trigger
Sample mode
Transmitting ECG data is being transmitted from the electrocardiograph to the PC
through the net or serial port in the AUTO or RHYT mode.
Transmit Fail ECG data fails to be transmitted from the electrocardiograph to the
PC through the net or serial port in the AUTO or RHYT mode.
Detecting The examining process of arrhythmia data in the Trigger Sample
mode
Memory Full The amount of files on the File Manage interface exceeds the
capacity limit. (SE-601A can accommodate 50 files, SE-601B can
accommodate 100 files, and SE-601C can accommodate 200 files.
For SE-601A/B, you can input special password to enable the storage
upgrade function, which can extend the storage to 200.)
Testing The period sampling is running.
Modu Error There is something wrong with the ECG board.
Demo The system is in the demonstration mode.
Overload The direct current offset voltage on an electrode is too high.
U Disk / USB Printer A U disk or a USB printer is connected to the USB interface.