SE-601 Series Electrocardiograph User Manual
- 14 -
2.5 Bottom Panel
Figure 2-6 SE-601 A / SE-601B / SE-601C Bottom Panel
Name Explanation
A Speaker Hole Path for sound from speaker (Only for SE-601B/C)
B Battery Compartment Compartment for lithium battery
C Heat Emission Hole Path for internal heat emission
D Label Position for product information label
E Fuse The fuse specification is T1AL250VP Ø5×20.
1) Battery Compartment
The rated voltage and the rated capacity of the rechargeable lithium battery are as follows:
Rated Voltage: 14.8V; Rated Capacity: 2200mAh.
WARNING :
1. Improper operation may cause the battery to be hot, ignited or exploded, and it
may lead to the decrease of the battery capacity. Therefore, it is necessary to
read the user manual carefully and pay more attention to warning messages.
2. When leakage or foul smell is found, stop using the battery immediately. If your
skin or cloth comes into contact with the leakage liquid, cleanse it with clean
water at once. If the leakage liquid splashes into your eyes, do not wipe them.
Irrigate them with clean water first and go to see a doctor immediately.
3. Only qualified service engineers authorized by the manufacturer can open the
battery compartment and replace the battery, and batteries of the same model
A
B
C
D
C
E