SE-601 Series Electrocardiograph User Manual
- 41 -
Sequence item, and then press F3 or F4 to select a recording sequence.
For details of the recording sequence, please see Section 9.1.6, “Specifying Recording
Sequence”.
5) If you set Sample Mode to 6×2 adjust, press Tab or Shift + Tab to move the cursor to
the Sample Time textbox, and then enter the sampling time within 10-24s.
6) Press Enter to confirm.
SE-601B / SE-601C Work Mode Setup Interface
SE-601A Work Mode Setup Interface
7) On the System Setup interface, press F1, F2, F3, F4, Tab or Shift + Tab to move the
cursor. When the cursor is on Lead, press Enter to open the Lead Setup interface.
8) For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Rhythm
Lead1/2/3 item, and then press F1 or F2 to select a lead.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Rhythm
Lead1/2/3 item, and then press F3 or F4 to select a lead.
9) For SE-601B/C, press Tab or Shift + Tab to move the cursor to the Lead Sequence
item, and then press F1 or F2 to select a sequence.
For SE-601A, press F1, F2, Tab or Shift + Tab to move the cursor to the Lead
Sequence item, and then press F3 or F4 to select a sequence.
10) Press Enter to confirm.