How to use
47
Reset to factory settings
Unstable ambient conditions (adapt filter)
Hardware (handshake mode)
Automatic without stability (printing mode)
*)
This is recommended step to do at the beginning of balance programming.
Internal balances are represented by either balance or weigh cell. In last case, programming
factors is not possible, unless external keyboard is available.
It is possible to adjust balance stability without reprogramming by using Stability Configuration
setting.
Balance calibration
Calibration of the balance is essential to ensure that correct weight is read.
To perform balance calibration:
1. Ensure that software is in the weighing mode: furnace is cold, Crucibles window buttons
to add crucibles are visible, Flow window is waiting for user action (waiting for a button to
be pressed).
Alternatively, as Administrator, it is possible to perform calibration at some other moment: pause
the flow, manually open furnace, etc.
2. Remove lid and crucibles carousels.
3. Replace massive ceramic balance pedestal with light-weighted plastic one.
Depending on the instrument, this step may be skipped or even balance shaft without the
pedestal used.
4. Press Tare button in the Balance window.
5. Press Calibrate button.
6. Wait for the message PLACE CALIBRATION WEIGHT.
7. Place calibration weight on the balance pedestal.
8. Wait until message CALIBRATING will disappears.
9. Return instrument into normal conditions: put balance pedestal and carousels back.
Do not forget to tare balance before un-pausing the flow as administrator.
Balance statuses:
Receiving blank text from the balance