How to use
55
If crucible is not empty (contains sample), then deleting this crucible required selection and
confirmation.
Pos: 9.4.14 /00 25 Überschriften /ELTRA 1.1 Übersc hriften/11 E xport @ 7\mod _13839003353 44_9.docx @ 5 1913 @ 2 @ 1
7.7 Export
Pos: 9.4.15 /00 10 ELTRA/0005 so ftware_hel p/Tga/How to use /Export @ 7\ mod_1383900 308984_9.doc x @ 51902 @ @ 1
After analysis is finished Results are saved. They can be further exported, if export is enabled.
To enable export:
1 Create export definition file.
2. Assign path to this file as Export for an Application.
3. Data will be exported at the end of analysis Phase.
Only existing crucibles data will be exported.
Export definition file is a special text file containing configuration for the export and definition
info: which data and how to export it. This file can be created from a scratch or, preferably,
existing LIMS\export.definition.txt file can be copied under a new name and modified.
Export definition file consists of lines where any line can be either empty, a comment (begins
from semicolon) or contain a definition.
Definitions
Definitions are case-insensitive (unless specified otherwise) and their format is
PARAMETER=VALUE. Definition with default value can be omitted in case, if default value is
what is needed.
DESTINATION=FILE
Configures LIMS export to work with the file.
Mandatory definition and should be always present.
PATH=path
Specify full path of the exporting file. Path can contain {0} to use current date/time and {1} to
use FILENAME parameter from the Import. See http://msdn.microsoft.com/en-
us/library/26etazsy(VS.80).aspx for information about formatting.
OPENWITH=open
Specify external program used to open exported file. Can be an exe-file, batch-file or windows
alias (to example, Notepad ). The program first parameter will be a full path of exported file.
Default (blank do not open with): OPENWITH=
TEXT=text
Output specified text into exported file. Text can contain character escapes to insert special
characters. See http://msdn.microsoft.com/en-us/library/4edbef7e.aspx for more information.
VAR=value
Insert specified value into exported file.