Short Manual Biometra TRIO
102
13 Short Manual
WARNING
Before initial operation check the correct operating voltage is set on the voltage selec-
tion switch on the underside of the device. Do not open the housing
! Danger of electric
shock!
CAUTION
The sample
block and the heated lid can reach high temperatures during operation.
Both the sample block and the heated lid can burn you. The samples are heated up
quickly and may come to an explosive boil. Wear goggles when handling hot samples.
Make sure that the li
d is closed securely before starting the program!
NOTICE
The use of oil between samples and sample block is not necessary. If you wish to use oil,
never use silicone oil. Use mineral oil.
Ensure that the ventilation slits are clear.
Important: Release
the lid wheel before opening the lid! Adjust the lid wheel for
every individual run!
After the Biometra TRIO has passed the power-on self-test the login screen is displayed:
Press Login to log in as existing user
Press Quick start for the non-user specific quick start of programs
Press EN/DE/CN to set the login screen language
After log in the Biometra TRIO home screen opens:
Log in
Home Screen