Biometra TRIO Maintenance and Repair
109
15 Maintenance and Repair
15.1 Factory calibration
This instrument is supplied fully calibrated by the manufacturer before delivery, using
equipment traceable to national standards. The Biometra Quality Management-System
is certified by BSI for BS EN ISO 9001:2015 and the used measurement system is tested
by an accredited testing laboratory in accordance with EN ISO 17025 at prescribed inter-
vals and gets regularly calibrated. If the instrument is installed in an environment that
meets the requested working conditions (see section "Technical specifications" p. 114), a
temperature verification or calibration after installation is not required.
15.2 Servicing and Repair
The Biometra TRIO is mainly maintenance-free. The care and maintenance tasks which
can be performed by the customer are limited to the cleaning and disinfection of the
housing and sample block.
All maintenance work and repairs beyond this scope must only be performed by Bi-
ometra GmbH customer service personnel or authorized and trained persons. Any unau-
thorized intervention limits warranty entitlements. If the device exhibits any faults or
defects, please contact the Biometra GmbH customer service department immediately.
In order to guarantee sound and safe operation and to ensure laboratory certification,
we recommend concluding a maintenance agreement with regular device validation.
15.3 Cleaning
WARNING
Risk of electric shock! Prior to commencing any maintenance or cleaning work, switch
off the device and pull out the mains plug. After cleaning, wait until the Biometra
TRIO
is completely dry before recommissioning it.
WARNING
If you are handling infectious and pathogenic materials observe the safety regulations.
Decontaminate the device in accordance with section "
Disinfecting the device" p. 110.
The professional decontamination of radioactive contaminations depends on the type of
substance used. Please contact your radiation protection off
icer on this matter and ob-
serve the relevant safety regulations!
NOTICE
The use of alcohol, organic solvents or abrasives can damage the paintwork.