EasyManua.ls Logo

Endress+Hauser Analytik Jena Biometra TRIO - Firmware Update and Spare Parts

Endress+Hauser Analytik Jena Biometra TRIO
118 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Maintenance and Repair Biometra TRIO
110
Only wipe the housing of the Biometra TRIO with a soft clean cloth which may be
slightly wetted with a commercial neutral detergent, if necessary.
Clean the ventilation slits on the underside and at the rear of the device with a vac-
uum cleaner.
15.4 Disinfecting the device
WARNING
Biological hazard! Clean the Biometra
TRIO with particular care after analyzing poten-
tially infectious material. Wear suitable protective equipment, e.g., protective gloves.
NOTICE
The only suitable cleaning method for the housing is wipe disinfection.
When using spray disinfectants there is a risk that the liquid enters the sensitive elec-
tronic system through the ventilation slits. If the disinfectant has a spray nozzle, apply
disinfectant to a suitable cloth before using it on the device.
Avoid contamination by handling sample substances with care.
Wipe spilled samples or reagents immediately with an absorbent cloth or piece of
paper.
If the Biometra TRIO is used for the analysis of infectious material, great care must
be taken because the TRIO cannot be decontaminated as a whole device.
Immediately remove visible contamination by suitable means. Do not allow solvents
to enter the device.
The sample block is also suitable for wipe and spray disinfection. The only suitable
cleaning method for the housing is wipe disinfection.
Device part Recommended disinfectants Provider
Sample chamber Descosept AF
Meliseptol HBV (cloths)
Dr. Schuhmacher GmbH
B. Braun
Housing Descosept Spezial Dr. Schuhmacher GmbH
15.5 Firmware update
For instruction for firmware upgrade, please contact the Biometra service department or
your local distributor/sales representative.
15.6 Spare parts
Only original spare parts mentioned in these operating instructions are allowed.

Table of Contents

Other manuals for Endress+Hauser Analytik Jena Biometra TRIO

Related product manuals