Proline Promag 55 PROFIBUS DP/PA Commissioning
Endress+Hauser 113
Example (change the output byte)
Output data
EMPTY_MODULE module
The measuring device is a so-called modular PROFIBUS slave. In contrast to a compact slave, the
structure of a modular slave is variable - it consists of several individual modules. In the GSD file,
the individual modules are described with their individual properties. The modules are permanently
assigned to the slots, i.e. the sequence or arrangement of the modules must be observed when
configuring the modules. Gaps between configured modules have to be assigned the
EMPTY_MODULE module.
For a more detailed description ä 96.
From To Result
0 2 Positive zero return is switched on.
2 0No effect
0 3 Positive zero return is switched off.
3 2No effect
Byte 1
Control

Table of Contents