Installation Proline Promag 55 PROFIBUS DP/PA
16 Endress+Hauser
3.2.4 Inlet and outlet runs
If possible, install the sensor well clear of fittings such as valves, T-pieces, elbows, etc.
Compliance with the following requirements for the inlet and outlet runs is necessary in order to
ensure measuring accuracy.
Inlet run 5 × DN
•Outlet run 2 × DN
A0003210
Fig. 12: Inlet and outlet runs
3.2.5 Vibrations
Secure the piping and the sensor if vibration is severe.
"
Caution!
It is advisable to install the sensor and transmitter separately if vibration is excessively severe.
Information on the permitted resistance to vibration and shock ä 145.
A0003208
Fig. 13: Measures to prevent vibration of the measuring device (L > 10 m / 33 ft)
5xDN
2xDN
L